Skip to content
PROGIENE, Protection & Hygiène
Actosed HP Ready 5 L jerrycan, with a tube of test strips and the activator bottle
InstrumentsHigh-level disinfection

Actosed HP Ready

High-level disinfectant based on peracetic acid.

Solution for the final disinfection by immersion of pre-cleaned, reusable, heat-sensitive semi-critical and critical medical devices. Biodegradable, aldehyde-free.

Packaging and references
1 format

Overview

Applications

  • Cleaned flexible and rigid endoscopes
  • Medical and surgical instruments
  • Manual disinfection bath or automatic endoscope washer

Key benefits

  • Combination of peracetic acid and hydrogen peroxide
  • Broad spectrum, including sporicidal, in 5 minutes
  • Activated solution stable for 5 days
  • Can be disposed of without neutralisation according to the manufacturer

Technical characteristics

Composition

Solution (4750 ml jerrycan): stabiliser, corrosion inhibitor and excipients. Activator (250 ml): ≤ 6% peracetic acid. The activated product initially contains ≥ 2000 ppm peracetic acid and ≥ 5000 ppm hydrogen peroxide.

Material compatibility

Steel, glass, ceramics, silicone and plastics.

Efficacy and standards

  • Bactericidal5 min

    EN 14561

    Ready to use

  • Fungicidal5 min

    C. albicans, A. brasiliensis

    EN 14562

    Ready to use

  • Tuberculocidal, mycobactericidal5 min

    EN 14563

    Ready to use

  • Virucidal5 min

    Polio, Adeno, MNV

    EN 14476

    Ready to use

  • Sporicidal5 min

    Bacillus subtilis

    EN 17126

    Ready to use

Instructions for use

Dilution or dose
Activate before use: pour the whole activator bottle (250 ml) into the jerrycan (4750 ml), close and mix. Ready to use, no dilution.
Contact time
5 min
  1. Pre-clean instruments with an enzymatic cleaner (the manufacturer recommends Actoclean Perfect), rinse surfaces and channels and dry.
  2. Immerse the instruments so that all surfaces and cavities are in contact; keep the bath closed.
  3. After the contact time, rinse with sterile water or water of at least drinking quality for at least 1 minute, then dry. Do not reuse rinse water.
  4. Check the solution with test strips; the prepared solution is stable for 5 days.
  5. Wear personal protective equipment (gloves, goggles).

Packaging and references

FormatReference
5 L (4750 ml + 250 ml activator)01.7303.5